Recent innovations made by Lightguide result in our best laser fiber for soft tissue laser surgery. The broad cylindrical 360° laser light emission of our Infinity Side Fiber ensures smooth and gentle laser energy delivery to the target tissue. Intravascular, intrastitial and intracutaneous laser ablation treatments can benefit from the smoother procedure and higher patient comfort.
The optimized fused fiber tip design guarantees first-class safety and durability. The optical fiber has marking lines to ease positioning in the target tissue. A compatible Introducer Set for vascular access is available.
Options upon request:
The fiber optic probes for radial emission applications that we produce are optimized for use @1470 nm. This product is particularly suitable for phlebology and gynecology applications where a radial emission of the laser light is recommended.
FIBERS
Our fibers are USP Class VI certified, non-toxic, and biocompatible. Standard products use either a 400 µm or a 600 µm core all silica fiber, to suit different types of lasers and medical procedures.
CONNECTOR
The probe is equipped with the Powershot SMA connector manufactured by SEDI-ATI. This connector is specially designed for power applications of disposable probes where cost issue is important. A free-standing ferrule with an air gap around the fiber allows excellent power handling and the assembly technique avoids any stress on the fiber itself.
DISTAL END
The most efficient design is obtained by a specific shaping of the fiber end which is enclosed in a silica capillary. The result is an annular emission of the laser energy for homogenous action on the vein wall.
JACKET
Ring fibers with a capillary assembly at the distal end are jacketed with a biocompatible material. This jacket has a cm marking to help visualize the fiber introduction in the vein.
MANUFACTURING
QUALIFICATION
Internal Qualification Reports (RQ150511 for RING fiber 400 µm and RQ150703 for RING fiber 600 µm): power handling of 400 Joules per cm on a tissue length of 80 cm @ 1470 nm.
Destructive test with 5000 J in the same volume of tissue creating a big hole + carbonization layer on the surface tip but not disconnecting the capsule.
Traction of the fiber tip of 10 Kgs.
The Bare Fiber is available in several different product variations to perfectly meet a wide variety of requirements and applications. There are different materials and geometries to choose from: from thin and extremely flexible to rigid and durable.
You may choose between the classic flat tip design or alternative shapes like a ball tip, conical tip, spherical tip and more. An integrated handpiece is also available. By catering to a patient’s anatomy on a case-dependent basis, access to the treatment site can be improved and the laser’s energy can be transmitted most efficiently.
Many variations of our Bare Fiber are available, and multiuse versions exist as well. Steam autoclave reconditioning helps reduce expenses on consumables for each treatment session.
Complementary with the Bare Fiber, the Dual LuerLock Handpiece can be used for precise alignment of the fiber tip to the tissue that needs to be treated.